産品中(zhōng)心

人嗜酸粒細胞趨化因子(ECF)酶聯免疫吸附檢測試劑盒
ELK1009
規格: 價格:
48T ¥1680.00
96T ¥2400.00

Overview

Product name: Human ECF(Eosinophil Chemotactic Factor) ELISA Kit
Reactivity: Human
Alternative Names: CCL11; SCYA11; Eotaxin 1; Eosinophil Chemotactic Protein 1; Chemokine C-C-Motif Ligand 11; Small Inducible Cytokine Subfamily A(Cys-Cys)Member 11
Assay Type: Sandwich
Sensitivity: 6.4 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Cytokine;Infection immunity;Autoimmunity;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human ECF. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human ECF. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human ECF, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human ECF in the samples is then determined by comparing the OD of the samples to the standard curve.

标準曲線

Concentration (pg/mL) OD Corrected OD
1000.00 2.401 2.276
500.00 1.911 1.786
250.00 1.316 1.191
125.00 0.831 0.706
62.50 0.453 0.328
31.25 0.304 0.179
15.63 0.230 0.105
0.00 0.125 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant ECF and the recovery rates were calculated by comparing the measured value to the expected amount of ECF in samples.
Matrix Recovery range Average
serum(n=5) 86-98% 92%
EDTA plasma(n=5) 95-107% 101%
Heparin plasma(n=5) 83-95% 89%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ECF and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 82-95% 89-103% 79-91% 85-93%
EDTA plasma(n=5) 87-103% 79-92% 86-99% 84-97%
Heparin plasma(n=5) 85-98% 97-104% 92-101% 82-96%
關閉

在線咨詢

Online consultation

  • 在線咨詢
  • 技術支持

關注微信公衆号

微信掃一(yī)掃立即咨詢

微信掃一(yī)掃立即咨詢

關閉