人生(shēng)長調節緻癌基因a(GROa/CXCL1)酶聯免疫吸附檢測試劑盒
ELK1024
規格: 價格:
48T ¥1680.00
96T ¥2400.00

Overview 文獻

Product name: Human GROa/CXCL1(Growth Regulated Oncogene Alpha) ELISA Kit
Reactivity: Human
Alternative Names: CXCL1; GRO1; GROa; GRO-A; MGSA-A; NAP3; SCYB1; Chemokine C-X-C-Motif Ligand 1; Melanoma Growth Stimulating Activity Alpha; Fibroblast Secretory Protein; Neutrophil Activating Protein 3
Assay Type: Sandwich
Sensitivity: 6.1 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Cytokine;Tumor immunity;Infection immunity;
Uniprot ID: P09341
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human GROa/CXCL1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human GROa/CXCL1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human GROa/CXCL1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human GROa/CXCL1 in the samples is then determined by comparing the OD of the samples to the standard curve.

标準曲線

Concentration (pg/mL) OD Corrected OD
1000.00 2.168 2.027
500.00 1.841 1.700
250.00 1.384 1.243
125.00 0.969 0.828
62.50 0.649 0.508
31.25 0.481 0.340
15.63 0.315 0.174
0.00 0.141 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant GROa/CXCL1 and the recovery rates were calculated by comparing the measured value to the expected amount of GROa/CXCL1 in samples.
Matrix Recovery range Average
serum(n=5) 82-97% 89%
EDTA plasma(n=5) 95-107% 101%
Heparin plasma(n=5) 81-95% 88%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of GROa/CXCL1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-101% 78-96% 79-98% 93-102%
EDTA plasma(n=5) 88-97% 91-103% 86-92% 89-97%
Heparin plasma(n=5) 88-104% 87-96% 85-93% 82-90%