産品中(zhōng)心

小(xiǎo)鼠半胱氨酸蛋白(bái)酶抑制劑3(CST3)酶聯免疫吸附檢測試劑盒
ELK1046
規格: 價格:
48T ¥1680.00
96T ¥2400.00

Overview

Product name: Mouse CST3(Cystatin 3) ELISA Kit
Reactivity: Mouse
Alternative Names: Cys-C; Cystatin C; Gamma Trace; Post-Gamma-Globulin; Neuroendocrine Basic Polypeptide; Amyloid Angiopathy And Cerebral Hemorrhage
Assay Type: Sandwich
Sensitivity: 0.126 ng/mL
Standard: 20 ng/mL
Detection Range: 0.32-20 ng/mL
Sample Type: Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids.
Assay Length: 3.5h
Research Area: Kidney biomarker;Urology;
Uniprot ID: P21460
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse CST3. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse CST3. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse CST3, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse CST3 in the samples is then determined by comparing the OD of the samples to the standard curve.

标準曲線

Concentration (ng/mL) OD Corrected OD
20.00 2.082 1.895
10.00 1.761 1.574
5.00 1.251 1.064
2.50 0.862 0.675
1.25 0.503 0.316
0.63 0.401 0.214
0.32 0.224 0.037
0.00 0.187 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant CST3 and the recovery rates were calculated by comparing the measured value to the expected amount of CST3 in samples.
Matrix Recovery range Average
serum(n=5) 78-90% 84%
EDTA plasma(n=5) 89-107% 98%
Heparin plasma(n=5) 87-99% 93%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CST3 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-92% 79-96% 87-103% 90-99%
EDTA plasma(n=5) 88-97% 96-105% 91-103% 89-97%
Heparin plasma(n=5) 85-94% 87-96% 79-95% 95-102%
關閉

在線咨詢

Online consultation

  • 在線咨詢
  • 技術支持

關注微信公衆号

微信掃一(yī)掃立即咨詢

微信掃一(yī)掃立即咨詢

關閉